Agilent Technologies Signal Analyzer N9030a User Manual
Page 992
6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Recall
Couplings
When a limit line is loaded from mass storage, it is automatically turned on. This allows the user
to see it, thus confirming the load. The Margin settings will match those when the limit was
saved
The instrument cannot mix Limits domains (X Axis Unit must be Frequency or Time for both
Limits). So when a Limits file is loaded, the analyzer will set the Limits domain (X Axis Unit) to
match that of the file. If this changes the Limits domain from what it was before the file was
loaded, all Limits data in all Limits sets will be erased before the data loads. If this operation is
over the remote interface there will be no warning if this occurs, so care should be taken to know
the domain of the file you are loading.
Readback
Selected Limit Line
Status Bits/OPC
dependencies
Sequential - aborts the current measurement
Initial S/W Revision
A.02.00
Limit Selection
These keys let you pick which Limit Line to import the data into; either 1, 2, 3, 4, 5 or 6. The
default is the currently selected limit. Once selected, the key returns back to the Import Data
menu and the selected Limit Line number is annotated on the key. Now you have selected exactly
what needs to be imported. In order to trigger an import of the selected Limit Line, you must
select the Open key in the Import Data menu.
An example of using this menu is: If you select 2 and continue to the File Open menu, Limit Line
2 will be imported from the file selected or entered in File Name option in the File Open dialog.
Key Path
Recall, Data, Limit Line
Notes
Auto return
Preset
Not part of Preset, but is reset to LLINE1 by Restore Mode Defaults; survives a shutdown.
State Saved
The selected limit number is saved in instrument state
Initial S/W Revision
A.02.00
Limit Selection
These keys let you pick which Limit Line to import the data into; either 1, 2, 3, 4, 5 or 6. The
default is the currently selected limit. Once selected, the key returns back to the Import Data
menu and the selected Limit Line number is annotated on the key. Now you have selected exactly
what needs to be imported. In order to trigger an import of the selected Limit Line, you must
select the Open key in the Import Data menu.
992
Remote Language Compatibility Measurement Application Reference