Agilent Technologies Signal Analyzer N9030a User Manual
Page 805
6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Meas Setup
Preset
Upper for Line 1, 3, and 5; Lower for Line 2, 4, 6.
Not affected by Mode Preset, preset by Restore Mode Defaults.
State Saved
Saved in instrument state.
Initial S/W Revision
A.02.00
Interpolation
Accesses a menu which lets you set the frequency and amplitude interpolation of the selected
limit.
Key Path
Meas Setup, Limits, Properties
Readback
[Lin|Log Frequency, Lin|Log Amplitude]
Initial S/W Revision
A.02.00
Frequency Interpolation
This key is grayed out if Time is the selected X Axis Units. Sets the interpolation between
frequency points, allowing you to determine how limit trace values are computed between points
in a limit table. The available interpolation modes are linear and logarithmic. If frequency
interpolation is logarithmic (Log), frequency values between limit points are computed by first
taking the logarithm of both the table values and the intermediate value. A linear interpolation is
then performed in this logarithmic frequency space. An exactly analogous manipulation is done
for logarithmic amplitude interpolation.
Note that the native representation of amplitude is in dB.
For linear amplitude interpolation and linear frequency interpolation, the interpolation is
computed as:
For linear amplitude interpolation and log frequency interpolation, the interpolation is computed
as:
For log amplitude interpolation and linear frequency interpolation, the interpolation is computed
as:
Remote Language Compatibility Measurement Application Reference