5  Legacy Command Descriptions

RB (Resolution Bandwidth)

bandwidths, and sweep time are normally coupled, but executing RB decouples them. Execute

"CR

(Couple Resolution Bandwidth)" on page 267

to re-establish coupling.

Format

RB HZ|KHZ|MHZ|GHZ|KZ|MZ|GZ

RB UP|DN

UP or DN increments in a 1, 3, 10 sequence

RB AUTO|MAN (8560 series only)

RB OA

RB?

Query Data Type

N/A

SCPI Equivalent Commands

[:SENSe]:BANDwidth[:RESolution]

[:SENSe]:BANDwidth[:RESolution]?

[:SENSe]:BANDwidth[:RESolution]:AUTO OFF|ON|0|1

[:SENSe]:BANDwidth[:RESolution]:AUTO?

(See

"Res BW " on page 549

)

Preset

8560 series: Coupled mode, 1 MHz

8566A/B, 8568A/B: Coupled mode, 3 MHz

Notes

Default values on X-Series instruments may vary from the legacy analyzers. Refer to the X-
Series User’s and Programmer’s Reference to find out any restrictions that may apply.

Remote Language Compatibility Measurement Application Reference

443