5 Legacy Command Descriptions
TDF (Trace Data Format)
TDF (Trace Data Format)
Syntax
Legacy Products
8560 series, 8566A/B, 8568A/B
Description
Formats trace information for return to the controller.
The different trace data formats are as follows:
Option
Format
P
Parameter data format. Numbers are in Hz, Volts, Watts, dBm, dBmV, DBuV, DBV.
A
Returns data as an A-block data field. MDS determines whether data comprises one or two 8-bit bytes. (See
"MDS (Measurement Data Size)" on page 374
.)
I
Returns data as an I-block data field. MDS determines whether data comprises one or two 8-bit bytes. (See
"MDS (Measurement Data Size)" on page 374
.)
M
ASCII data format.
B
Binary data format. MDS determines whether data comprises one or two 8-bit bytes. (See
.)
Format
TDF P|A|I|M|B
TDF?
Query Data Type
P|A|I|M|B
SCPI Equivalent Commands
:FORMat[:TRACe][:DATA] ASCii | INTeger,32 | REAL,32 | REAL,64
(See
"Format Data: Numeric Data (Remote Command Only)" on page 124
Preset
P
Remote Language Compatibility Measurement Application Reference