Agilent Technologies Signal Analyzer N9030a User Manual
Page 1403
6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Trigger
Trig Slope
Controls the trigger polarity. It is set positive to trigger on a rising edge and negative to trigger on
a falling edge.
Key Path
Trigger, Video
Remote Command
:TRIGger[:SEQuence]:VIDeo:SLOPe POSitive|NEGative
:TRIGger[:SEQuence]:VIDeo:SLOPe?
Example
TRIG:VID:SLOP NEG
Preset
POSitive
State Saved
Saved in instrument state
Backwards
Compatibility SCPI
:TRIGger[:SEQuence]:IF:SLOPe NEGative|POSitive
:TRIGger[:SEQuence]:IF:SLOPe?
For backward compatibility with VSA/PSA comms apps
Backwards Compatibility
Notes
The legacy :TRIGger[:SEQuence]:SLOPe command affects the slopes for the VID, LINE, EXT1,
EXT2, and RFB triggers.
Initial S/W Revision
Prior to A.02.00
Remote Command
:TRIGger[:SEQuence]:SLOPe POSitive|NEGative
:TRIGger[:SEQuence]:SLOPe?
Example
TRIG:SLOP NEG
Preset
POSitive
State Saved
Saved in instrument state
Backwards Compatibility
Notes
In ESA/PSA, the Trigger Slope was global to all triggers. In the X-Series, the slope can be set
individually for each Trigger Source. For backward compatibility, the global SLOPe command
updates all instances of trigger slope (VID, LINE, EXT1, EXT2, TV, RFB). The query returns the
trigger slope setting of the currently selected trigger source.
Initial S/W Revision
Prior to A.02.00
Trig Delay
Controls a time delay during that the analyzer will wait to begin a sweep after meeting the trigger
criteria. You can use negative delay to pre-trigger the instrument in the time domain or FFT, but
not in swept spans.
Key Path
Trigger, Video
Remote Language Compatibility Measurement Application Reference