Agilent Technologies Signal Analyzer N9030a User Manual
Page 170
3 Programming the Analyzer
STATus Subsystem
Questionable Power Negative Transition
This command determines which bits in the Questionable Power Condition register will set the
corresponding bit in the Questionable Power Event register when the condition register bit has a
negative transition (1 to 0).
The variable
Mode
All
Remote
Command
:STATus:QUEStionable:POWer:NTRansition
:STATus:QUEStionable:POWer:NTRansition?
Example
STAT:QUES:POW:NTR 32
50 MHz Input Power became OK for Cal will be reported to the Power Summary of the Status Questionable
register.
Preset
0
Min
0
Max
32767
Status
Bits/OPC
dependencies
Sequential command
Initial S/W
Revision
Prior to A.02.00
Questionable Power Positive Transition
This command determines which bits in the Questionable Power Condition register will set the
corresponding bit in the Questionable Power Event register when the condition register bit has a
positive transition (0 to 1).
The variable
Mode
All
Remote
Command
:STATus:QUEStionable:POWer:PTRansition
:STATus:QUEStionable:POWer:PTRansition?>
Example
STAT:QUES:POW:PTR 32
50 MHz Input Power became too high for Cal will be reported to the Power Summary of the Status
Questionable register.
Preset
32767
Min
0
170
Remote Language Compatibility Measurement Application Reference