Security – Agilent Technologies Signal Analyzer N9030a User Manual
Page 1312
6 RLC Swept SA Measurement Front-Panel & SCPI Reference
System
Notes
Return value is the host ID as a string
Initial S/W Revision
Prior to A.02.00
Security
Accesses capabilities for operating the instrument in a security controlled environment.
Key Path
System
Initial S/W Revision
A.04.00
USB
The Windows operating system can be configured to disable write access to the USB ports for
users who are in a secure environment where transferring data from the instrument is
prohibited. This user interface is a convenient way for the customer to disable write access to
USB.
Key Path
System, Security
Mode
All
Scope
Mode Global
Remote Command
:SYSTem:SECurity:USB:WPRotect[:ENABle] ON|OFF|0|1
:SYSTem:SECurity:USB:WPRotect[:ENABle]?
Example
:SYST:SEC:USB:WPR ON Will set USB ports to Read-only
Notes
When the USB ports are in Read-only mode then no data can be stored to USB, including the
internal USB memory used for a back-up location for the calibration data.
Dependencies
This key is grayed-out unless the current user has administrator privileges.
Preset
This is unaffected by Preset or any Restore System Defaults. An Agilent Recovery will set the
USB to write protect OFF
State Saved
No
Range
Read-Write|Read only
Initial S/W Revision
A.04.00
1312
Remote Language Compatibility Measurement Application Reference