6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Marker
Select Marker
Specifies the selected marker. The term “selected marker” is used throughout this document to
specify which marker will be affected by the functions.
Key Path
Marker
Notes
The selected marker is remembered even when not in the Marker menu and is used if a Search
is done or a Band Function is turned on or for Signal Track or Continuous Peak.
Preset
Marker 1
State Saved
The number of the selected marker is saved in instrument state.
Initial S/W Revision
Prior to A.02.00
Normal
Sets the control mode for the selected marker to Normal and turns on the active function for
setting its value. If the selected marker was Off, it is placed at the center of the screen on the trace
specified by the marker’s Trace attribute.
A Normal mode
(POSition type) marker can be moved to any point on the X Axis by specifying
its X Axis value. Its absolute Y Axis value is then the value of the trace point at that X Axis value.
Key Path
Marker
Example
:CALC:MARK:MODE POS sets Marker 1 to Normal.
Notes
See the description under the “ Marker” key.
Couplings
The marker addressed by this command becomes the selected marker on the front panel.
State Saved
The marker control mode (Normal, Delta, Fixed, Off) and X Axis value are saved in instrument
state.
Initial S/W Revision
Prior to A.02.00
Backwards
Compatibity SCPI
:CALCulate:MARKer[1]|2|...12:STATe ON|1
Setting a marker which is OFF to ON or 1 selects the marker, puts it in Normal mode and places it
at the center of the screen.
Setting a marker which is not OFF to ON has no effect (does not change its control mode).
Example: CALC:MARK2:STAT ON sets Marker 2 to Normal if it was off; otherwise it does nothing.
The response to the query will be ON unless the marker is OFF.
Remote Language Compatibility Measurement Application Reference