6 RLC Swept SA Measurement Front-Panel & SCPI Reference
System
Remote Command
:SYSTem:COMMunicate:LAN:SCPI:SOCKet:ENABle OFF|ON|0|1
:SYSTem:COMMunicate:LAN:SCPI:SOCKet:ENABle?
Example
:SYST:COMM:LAN:SCPI:SOCK:ENAB OFF
Preset
This is unaffected by a Preset but is set to ON with a “Restore System Defaults->Misc”
State Saved
No
Range
On | Off
Initial S/W Revision
Prior to A.02.00
SICL Server
Turns the SICL server capability On or Off, enabling you to limit SCPI access over LAN through
the SICL server. (SICL IEEE 488.2 protocol.)
Parameter
Description
Setting
Maximum Connections
The maximum number of connections that can be accessed
simultaneously
5
Instrument Name
The name (same as the remote SICL address) of your analyzer
inst0
Instrument Logical Unit
The unique integer assigned to your analyzer when using SICL
LAN
8
Emulated GPIB Name
The name (same as the remote SICL address) of the device
used when communicating with your analyzer
gpib7
Emulated GPIB Logical Unit
The unique integer assigned to your device when it is being
controlled using SICL LAN
8
Emulated GPIB Address
The emulated GPIB address assigned to your transmitter
tester when it is a SICL server (the same as your GPIB
address)
18
Key Path
System, I/O Config, SCPI LAN
Mode
All
Remote Command
:SYSTem:COMMunicate:LAN:SCPI:SICL:ENABle OFF|ON|0|1
:SYSTem:COMMunicate:LAN:SCPI:SICL:ENABle?
Example
:SYST:COMM:LAN:SCPI:SICL:ENAB OFF
Preset
This is unaffected by Preset, but is set to ON with a “Restore System Defaults->Misc”
Remote Language Compatibility Measurement Application Reference