6 RLC Swept SA Measurement Front-Panel & SCPI Reference
System
l
"Query the Mechanical Relay Cycle Count" on page 1315
l
"Query the Operating Temperature Extremes" on page 1315
l
"Query the Elapsed Time since First Power-On" on page 1316
Query the Mechanical Relay Cycle Count
Returns the count of mechanical relay cycles.
For N9038A, there are additional 2 Mechanical Relays, which are
Remote Command
:SYSTem:MRELay:COUNt?
Example
:SYST:MREL:COUN?
Notes
Query Only
The return value is a comma-separated list of the individual counts for each mechanical relay.
The position of the relays in the list is:
“
Atten>,<30dB Atten>,
Items in the list not pertaining to your particular hardware configuration return with the value –
999.
Initial S/W Revision
Prior to A.02.00
Modified at S/W
Revision
A.08.00
Query the Operating Temperature Extremes
Returns the low (or high) operating temperature extreme value. The value survives a power-cycle
and is the temperature extreme encountered since the value was reset by the factory or service
center.
Mode
All
Remote Command
:SYSTem:TEMPerature:LEXTreme?
Example
:SYST:TEMP:LEXT?
Notes
Value (in degrees Celsius) at which the lowest operating temperature has been recorded since
first power-up.
State Saved
No
Initial S/W Revision
Prior to A.02.00
Remote Language Compatibility Measurement Application Reference