6  RLC Swept SA Measurement Front-Panel & SCPI Reference

Sweep/Control

The key is also grayed out if Gate Control = Level.

Preset

461.6 us

WiMAX OFDMA: 50 us

GSM/EDGE: 200 us

State Saved

Saved in instrument state

Min

100 ns

Max

5 s

Backwards
Compatibility SCPI

[:SENSe]:SWEep:TIME:GATE:LENGth 

ESA compatibility

Initial S/W Revision

Prior to A.02.00

Method

This lets you choose one of the three different types of gating.

Not all types of gating are available for all measurements.

Key Path

Sweep/Control, Gate

Remote Command

[:SENSe]:SWEep:EGATe:METHod LO|VIDeo|FFT

[:SENSe]:SWEep:EGATe:METHod?

Example

SWE:EGAT:METH FFT

Preset

LO

State Saved

Saved in instrument state

Initial S/W Revision

Prior to A.02.00

LO

When Gate is set to On, the LO sweeps whenever the gate conditions as specified in the Gate
menu are satisfied by the signal at the Gate Source.

This form of gating is more sophisticated, and results in faster measurements. With Gated LO, the
analyzer only sweeps while the gate conditions are satisfied. This means that a sweep could take
place over several gate events. It would start when the gate signal goes true and stop when it goes
false, and then continue when it goes true again. But since the LO is sweeping as long as the gate
conditions are satisfied, the sweep typically finishes much more quickly than with Gated Video.

Remote Language Compatibility Measurement Application Reference