Agilent Technologies Signal Analyzer N9030a User Manual
Page 1203
6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Sweep/Control
The key is also grayed out if Gate Control = Level.
Preset
461.6 us
WiMAX OFDMA: 50 us
GSM/EDGE: 200 us
State Saved
Saved in instrument state
Min
100 ns
Max
5 s
Backwards
Compatibility SCPI
[:SENSe]:SWEep:TIME:GATE:LENGth
ESA compatibility
Initial S/W Revision
Prior to A.02.00
Method
This lets you choose one of the three different types of gating.
Not all types of gating are available for all measurements.
Key Path
Sweep/Control, Gate
Remote Command
[:SENSe]:SWEep:EGATe:METHod LO|VIDeo|FFT
[:SENSe]:SWEep:EGATe:METHod?
Example
SWE:EGAT:METH FFT
Preset
LO
State Saved
Saved in instrument state
Initial S/W Revision
Prior to A.02.00
LO
When Gate is set to On, the LO sweeps whenever the gate conditions as specified in the Gate
menu are satisfied by the signal at the Gate Source.
This form of gating is more sophisticated, and results in faster measurements. With Gated LO, the
analyzer only sweeps while the gate conditions are satisfied. This means that a sweep could take
place over several gate events. It would start when the gate signal goes true and stop when it goes
false, and then continue when it goes true again. But since the LO is sweeping as long as the gate
conditions are satisfied, the sweep typically finishes much more quickly than with Gated Video.
Remote Language Compatibility Measurement Application Reference
1203