Ksy (video trigger), Syntax, Legacy products – Agilent Technologies Signal Analyzer N9030a User Manual
Page 350: Description, Syntax legacy products
5 Legacy Command Descriptions
KSy (Video Trigger)
KSy (Video Trigger)
Syntax
Legacy Products
8566A/B, 8568A/B
Description
Activates the normal video trigger mode. When KSy is executed, the RF input signal is only
displayed when the video trigger signal, which is internally triggered off the input signal, exceeds
the trigger threshold level.
Format
KSy
Query Data Type
N/A
SCPI Equivalent Commands
None
Notes
The functions of KSy are identical to the TM VID form of
and to
"T4 [four] (Video Trigger)" on page 480
If an 8566A/B or an 8568A/B analyzer is in zero span and the sweep time is less than 20
msec, the display is refreshed only when a fresh trace has been taken. This can cause the
displayed trace to flicker.
In X-Series instruments, all traces are displayed continuously, so are therefore free of
flicker.
350
Remote Language Compatibility Measurement Application Reference