Next peak, Next pk right – Agilent Technologies Signal Analyzer N9030a User Manual
Page 896
6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Peak Search
many times as required to navigate back through the previously accessed menus until you get
back to the Peak Search menu.
Next Peak
Pressing Next Peak moves the selected marker to the peak that has the next highest amplitude less
than the marker’s current value. Only peaks which meet all enabled peak criteria are considered.
If there is no valid peak lower than the current marker position, a “No peak found” message is
generated and the marker is not moved.
If the selected marker was off, then it is turned on as a normal marker and a peak search is
performed.
Key Path
Peak Search
Remote Command
:CALCulate:MARKer[1]|2|...12:MAXimum:NEXT
Example
CALC:MARK2:MAX:NEXT Selects marker 2 and moves it to the peak that is closest in amplitude
to the current peak, but the next lower value.
Notes
Sending this command selects the subopcoded marker.
State Saved
Not part of saved state.
Initial S/W Revision
Prior to A.02.00
Next Pk Right
Pressing Next Pk Right moves the selected marker to the nearest peak right of the current marker
that meets all enabled peak criteria. If there is no valid peak to the right of the current marker
position, a “No peak found” message is generated and the marker is not moved.
If the selected marker was off, then it is turned on as a normal marker and a peak search is
performed.
Key Path
Peak Search
Remote Command
:CALCulate:MARKer[1]|2|...12:MAXimum:RIGHt
Example
CALC:MARK2:MAX:RIGH Selects marker 2 and moves it to the next peak to the right of the
current marker position.
Notes
Sending this command selects the subopcoded marker.
State Saved
Not part of saved state.
Initial S/W Revision
Prior to A.02.00
896
Remote Language Compatibility Measurement Application Reference