3 Programming the Analyzer
STATus Subsystem
"Questionable Calibration Extended Failure Positive Transition" on page 154
Questionable Calibration Extended Failure Condition
This query returns the decimal value of the sum of the bits in the Questionable Calibration
Extended Failure Condition register.
The data in this register is continuously updated and reflects the current conditions.
Mode
All
Remote Command
:STATus:QUEStionable:CALibration:EXTended:FAILure:CONDition?
Example
STAT:QUES:CAL:EXT:FAIL:COND?
Preset
0
Status Bits/OPC dependencies
Sequential command
Initial S/W Revision
Prior to A.02.00
Questionable Calibration Extended Failure Enable
This command determines which bits in the Questionable Calibration Extended Failure Condition
Register will set bits in the Questionable Calibration Extended Failure Event register, which also
sets bit 9 of the Questionable Calibration Register. The variable
decimal values of the bits you want to enable.
Mode
All
Remote Command
:STATus:QUEStionable:CALibration:EXTended:FAILure:ENABle
:STATus:QUEStionable:CALibration:EXTended:FAILure:ENABle?
Example
STAT:QUES:CAL:EXT:FAIL:ENAB 1
Can be used to query if an EMI conducted alignment is needed.
Preset
32767
Min
0
Max
32767
Status Bits/OPC
dependencies
Sequential command
Initial S/W
Revision
Prior to A.02.00
152
Remote Language Compatibility Measurement Application Reference