6 RLC Swept SA Measurement Front-Panel & SCPI Reference
AMPTD Y Scale
configuration uses a single attenuator with combined mechanical and electronic sections that
controls all the attenuation functions. Different models in the X-Series come with different
configurations.
l
See
"Dual Attenuator Configurations" on page 518
l
See
"Single Attenuator Configuration" on page 519
Most Attenuation settings are the same for all measurements – they do not change as you change
measurements. Settings like these are called “Meas Global” and are unaffected by Meas Preset.
Key Path
AMPTD Y Scale
Scope
Meas Global
Dependencies
In measurements that support the I/Q inputs, this key is unavailable when I/Q is the selected
input, and is replaced by the Range key in that case.
Readback Line
Contains a summary in [ ] brackets of the current total attenuation. See the descriptions of the
"Enable Elec Atten" on page 521
, and
keys for more detail on the contributors to the total attenuation.
Note that when "Pre-Adjust for Min Clip" is on, this value can change at the start of every
measurement.
Initial S/W Revision
Prior to A.02.00
Modified at S/W
Revision
A.03.00
Dual Attenuator Configurations
Configuration 1: Mechanical attenuator + optional electronic attenuator
Configuration 2: Mechanical attenuator, no optional electronic attenuator