AMBPL (A minus B plus Display Line into A)
AMBPL (A minus B plus Display Line into A)
Syntax
Legacy Products
8560 series, 8566A/B, 8568A/B
Description
Does a point-by-point subtraction of Trace B from Trace A, and then adds the display line point
values to the difference. The results are sent to Trace A.
The query command AMBPL? returns different responses depending on the language being used.
Format
AMBPL (0|1|OFF|ON)
AMBPL?
Query Data Type
8560: 1 or 0, indicating ON or OFF state respectively.
8566A/B, 8568A/B: ON or OFF.
SCPI Equivalent Commands
None
Preset
OFF
Couplings
AMBPL
sets Trace B to View mode and turns AMB (Normalize) OFF. All trace math is
mutually exclusive, so turning one on turns the other off and vice versa. Similarly, when
AMBPL
is on and you change Trace B to Clearwrite or Maxhold, it turns AMBPL off.