Mktype (marker type), Syntax, Legacy products – Agilent Technologies Signal Analyzer N9030a User Manual
Page 410: Description, Syntax legacy products
5 Legacy Command Descriptions
MKTYPE (Marker Type)
MKTYPE (Marker Type)
Syntax
Legacy Products
8566A/B, 8568A/B
Description
Specifies the type of marker.
Type
Function
Commands Used to Position Marker
PSN
Allows the marker to be positioned horizontally in display units
(default)
"MKP (Marker Position)" on page 397
"MKF (Marker Frequency)" on page 387
AMP
Allows the marker to be positioned according to amplitude
"MKA (Marker Amplitude)" on page 381
FIXED
Allows a marker to be placed at any fixed point on the display
"MKP (Marker Position)" on page 397
"MKF (Marker Frequency)" on page 387
"MKA (Marker Amplitude)" on page 381
Format
MKTYPE PSN|AMP|FIXED
MKTYPE?
Query Data Type
PSN|AMP|FIXED
SCPI Equivalent Commands
:CALCulate:MARKer#:MODE POSition
:CALCulate:MARKer#:MODE FIXed
:CALCulate:MARKer#:X
:CALCulate:MARKer#:Y
(See
)
Preset
PSN
Notes
Marker type can only be set for an active marker. The marker type is reset to PSN when
the marker is turned off (using
"MKOFF (Marker Off)" on page 396
), or when the
instrument is preset.
410