Meas preset – Agilent Technologies Signal Analyzer N9030a User Manual
Page 859
6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Meas Setup
Couplings
If an SNS is connected, and the Type is set to SNS, this parameter turns the SNS on and off.
When an SNS is not connected this parameter turns the BNC 28V output on and off.
When the SA mode is first entered this parameter is set to OFF and the 28v drive turned OFF.
When the SA mode is exited this parameter is set to OFF and the 28v drive turned OFF.
Preset
OFF
State Saved
Saved in instrument state.
Backwards Compatibility
Notes
In previous Noise Figure analysis applications, this command
could optionally be preceded with the :SENSe keyword. The
optional :SENSe keyword is no longer supported.
Initial S/W Revision
Prior to A.02.00
SNS Attached (Remote Command Only)
If an Smart Noise Source (SNS) is present this command will return 1 otherwise it will return 0.
Remote Command
:SOURce:NOISe:SNS:ATTached?
Example
SOUR:NOIS:SNS:ATT?
Preset
OFF
State Saved
No
Backwards Compatibility
Notes
In previous Noise Figure analysis applications, this command
could optionally be preceded with the :SENSe keyword. The
optional :SENSe keyword is no longer supported.
Initial S/W Revision
Prior to A.02.00
Meas Preset
This key returns the Meas Local variables in the Swept SA measurement to their preset values.
This is the same as sending the SCPI command CONF:SAN.
The only exception is Limits On/Off, which is a persistent Meas Local variable. It will be set to Off
by a Mode Preset but not by Meas Preset.
Key Path
Meas Setup
Initial S/W Revision
Prior to A.02.00
Remote Language Compatibility Measurement Application Reference
859