6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Source
Mode
SA
Scope
Mode Global
Remote Command
:SOURce:FREQuency:OFFSet
:SOURce:FREQuency:OFFSet?
:SOURce:FREQuency:OFFSet:STATe ON|OFF|1|0
:SOURce:FREQuency:OFFSet:STATe?
Example
:SOUR:FREQ:OFFS 10MHz
Sets the source frequency offset to 10MHz.
Dependencies
If the currently selected source does not support this capability (for example, an internal
Tracking Generator which must track the LO), this key is forced to its Preset value and grayed
out
Preset
This is unaffected by Mode Preset but is set to 0.00Hz on a Source Preset or Restore
Input/Output Defaults.
State Saved
Part of the Input/Output system, which means it is Loaded and Saved with state
Min
–10 GHz
Max
10 GHz
Backwards
Compatibility SCPI
:SOURce:EXTernal:SWEep:OFFSet:FREQuency
:SOURce:EXTernal:SWEep:OFFSet:FREQuency?
The PSA option 215 supports this SCPI command. This command is equivalent to :
SOURce:FREQuency:OFFSet
Initial S/W Revision
A.06.01
Modified at S/W
Revision
A.10.01
Source Mode
Accesses the Source Mode softkey menu. This menu lets you select Tracking mode for the Source,
and also allows you to set the Source Mode to OFF.
The Source Mode can be set to Tracking without the user setting it directly. There are several
couplings that cause Source Mode to be automatically set to Tracking (detailed in the table
below). One important coupling is that Source Mode is forced to Tracking when the RF Output is
turned on if the measurement supports Tracking. Since Source Mode is set to Off on a Mode
Preset, this means that you will rarely need to change the Source Mode setting directly.
Note:
Remote Language Compatibility Measurement Application Reference