3 Programming the Analyzer
STATus Subsystem
Preset
32767
Min
0
Max
32767
Status Bits/OPC
dependencies
Sequential command
Initial S/W
Revision
Prior to A.02.00
Questionable Frequency Register
The following commands and queries are available for this register:
"Questionable Frequency Condition" on page 157
"Questionable Frequency Enable" on page 157
"Questionable Frequency Event Query" on page 158
"Questionable Frequency Negative Transition" on page 158
"Questionable Frequency Positive Transition" on page 159
Questionable Frequency Condition
This query returns the decimal value of the sum of the bits in the Questionable Frequency
Condition register.
The data in this register is continuously updated and reflects the current conditions.
Mode
All
Remote Command
:STATus:QUEStionable:FREQuency:CONDition?
Example
STAT:QUES:FREQ:COND?
Preset
0
Status Bits/OPC dependencies
Sequential command
Initial S/W Revision
Prior to A.02.00
Questionable Frequency Enable
This command determines which bits in the Questionable Frequency Condition Register will set
bits in the Questionable Frequency Event register, which also sets the Frequency Summary bit (bit
Remote Language Compatibility Measurement Application Reference