6 RLC Swept SA Measurement Front-Panel & SCPI Reference
User Preset
The User Preset function does the following:
•
Aborts the currently running measurement.
•
Sets the mode State to the values defined by Save User Preset.
•
Makes the saved measurement for the currently running mode the active measurement.
•
Brings up the saved menu for the power-on mode.
•
Clears the input and output buffers.
•
Sets the Status Byte to 0.
Key Path
User Preset
Remote Command
:SYSTem:PRESet:USER
Example
:SYST:PRES:USER:SAVE:SYST:PRES:USER
Notes
:SYST:PRES:USER:SAVE is used to save the current state as the user preset state.
Clears all pending OPC bits. The Status Byte is set to 0.
Pressing the User Preset front-panel key while already in the User Preset menu will cause the
User Preset to get executed
Couplings
A user preset will cause the currently running measurement to be aborted and cause the saved
measurement to be active. Recalling a User Preset file has the same issues that recalling a Save
State file has. Some settings may need to be limited and therefore re-coupled, since the
capabilities of the mode may have changes when the User Preset file was last saved.
Initial S/W Revision
Prior to A.02.00
User Preset All Modes
Recalls all of the User Preset files for each mode, switches to the power-on mode, and activates the
saved measurement from the power-on mode User Preset file.
When the instrument is secured, all of the user preset files are converted back to their default user preset
files.
The User Preset function does the following:
•
Aborts the currently running measurement.
•
Switches the Mode to the power-on mode.
•
Restores the User Preset files for each mode.
•
Makes the saved measurement for the power-on mode the active measurement.
•
Brings up the saved menu for the power-on mode.
1440