PP (Preselector Peak)

PP (Preselector Peak)

Syntax

Legacy Products

8560 series, 8566A/B

Description

Optimizes preselector tracking to peak the amplitude of a signal at the active marker. If a marker
is not on the screen, PP places a marker at the highest signal level, and optimizes preselector
tracking at that frequency.

Format

PP

Query Data Type

N/A

SCPI Equivalent Commands

[:SENSe]:POWer[:RF]:PCENter (see

"Presel Center" on page 530

)

Notes

This command is only supported when the X-series instrument’s maximum frequency
limit is greater than 3.6 GHz. If the command is sent to an instrument with a maximum
frequency limit of 3.6 GHz or less, the command is not executed, and no error is generated.