6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Recall
TRACE6,
Example
:MMEM:LOAD:TRAC DATA TRACE2, "myTrace2.csv" Imports the 2nd trace from the file
myTrace2.csv in the current path. The default path is My Documents\SA\data\traces
Dependencies
For SA measurements, a trace cannot be recalled from a trace file that was exported with ALL
traces selected.
A trace cannot be imported if the number of trace points in the file do not match the number of
sweep points currently set for the measurement. If this happens, an error message is
generated.
Errors are reported if the file is empty or missing, or if the file type does not match, or if there is a
mismatch between the file type and the destination data type.
If any error occurs while trying to load a file manually (as opposed to during remote operation),
the analyzer returns to the Import Data menu and the File Open dialog goes away.
Couplings
When a trace is imported, Trace Update is always turned OFF for that trace and Trace Display is
always turned ON.
Readback
Selected Trace
Status Bits/OPC
dependencies
Sequential - aborts the current measurement.
Initial S/W Revision
Prior to A.02.00
Select Trace
These keys let you pick the Trace into which to import the data; either 1, 2, 3, 4, 5 or 6. The
default is the currently selected trace, selected in this menu or in the Trace/Det, Export Data,
Recall Trace, or Save Trace menus; except if you have chosen All then All remains chosen until
you specifically change it to a single trace.
Once selected, the key returns back to the Import Data menu and the selected Trace number is
annotated on the key. The next step is to select the Open key in the Import Data menu.
Key Path
Recall, Data, Trace
Notes
Auto return
Couplings
When you select the trace into which to import the data, it makes that trace the current trace, so
it displays on top of all of the other traces.
Preset
Not part of Preset, but is reset to TRACE1 by Restore Mode Defaults; survives shutdown
State Saved
The current trace number is saved in State
Initial S/W Revision
Prior to A.02.00
988
Remote Language Compatibility Measurement Application Reference