Sngls (single sweep), Syntax, Legacy products – Agilent Technologies Signal Analyzer N9030a User Manual
Page 463: Description, Sngls (single sweep)" on, Syntax legacy products
5 Legacy Command Descriptions
SNGLS (Single Sweep)
SNGLS (Single Sweep)
Syntax
Legacy Products
8560 series, 8566A/B, 8568A/B
Description
Sets the instrument to single-sweep mode. Each time
is sent, one
sweep taken as long as the trigger conditions are met.
Format
SNGLS
Query Data Type
N/A
SCPI Equivalent Commands
:INITiate:CONTinuous 0 (see
"Cont (Continuous Measurement/Sweep)" on page 563
Notes
The functions of SNGLS are identical to
"S2 [two] (Single Sweep)" on page 457
Remote Language Compatibility Measurement Application Reference