STATus Subsystem
"Operation Event Query" on page 142
"Operation Negative Transition" on page 142
"Operation Positive Transition" on page 143
Operation Condition Query
This query returns the decimal value of the sum of the bits in the Status Operation Condition
register.
The data in this register is continuously updated and reflects the current conditions.
Mode
All
Remote Command
:STATus:OPERation:CONDition?
Example
STAT:OPER:COND?
Preset
0
Status Bits/OPC dependencies
Sequential command
Initial S/W Revision
Prior to A.02.00
Operation Enable
This command determines which bits in the Operation Event register, will set the Operation
Status Summary bit (bit 7) in the Status Byte Register. The variable
decimal values of the bits you want to enable.
The preset condition is to have all bits in this enable register set to 0. To have any Operation Events
reported to the Status Byte Register, one or more bits need to be set to 1.
Mode
All
Remote Command
:STATus:OPERation:ENABle
:STATus:OPERation:ENABle?
Example
STAT:OPER:ENAB 1 Sets the register so that Align Now operation will be reported to the Status
Byte Register.
Preset
0
Min
0
Max
32767
Status Bits/OPC
Sequential command
Remote Language Compatibility Measurement Application Reference