3 Programming the Analyzer
STATus Subsystem
Questionable Integrity Register
The following commands and queries are available for this register:
"Questionable Integrity Condition" on page 160
"Questionable Integrity Enable" on page 160
"Questionable Integrity Event Query" on page 161
"Questionable Integrity Negative Transition" on page 161
"Questionable Integrity Positive Transition" on page 162
Questionable Integrity Condition
This query returns the decimal value of the sum of the bits in the Questionable Integrity
Condition register.
The data in this register is continuously updated and reflects the current conditions.
Mode
All
Remote Command
:STATus:QUEStionable:INTegrity:CONDition?
Example
STAT:QUES:INT:COND?
Preset
0
Status Bits/OPC dependencies
Sequential command
Initial S/W Revision
Prior to A.02.00
Questionable Integrity Enable
This command determines which bits in the Questionable Integrity Condition Register will set bits
in the Questionable Integrity Event register, which also sets the Integrity Summary bit (bit 9) in
the Questionable Register. The variable
want to enable.
Mode
All
Remote
Command
:STATus:QUEStionable:INTegrity:ENABle
:STATus:QUEStionable:INTegrity:ENABle?
Example
STAT:QUES:INT:ENAB 8
Measurement Uncalibrated Summary will be reported to the Integrity Summary of the Status Questionable
register.