Fixed marker z axis value – Agilent Technologies Signal Analyzer N9030a User Manual
Page 683
6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Marker
Example
:CALC:MARK:MODE FIX sets Marker 1 to Fixed.
Notes
See the description under the “ Marker” key, above.
Dependencies
•
You cannot directly set the X or Y value of a Fixed marker which has a marker function turned
on. If an attempt is made to actually adjust it while a Marker Function is on, a warning
message is generated.
•
You cannot directly set the Y value of a Fixed marker while Normalize is turned on. If an
attempt is made to do so while Normalize is on, a warning message is generated.
State Saved
The marker control mode (Normal, Delta, Fixed, Off) and X and Y Axis values are saved in
instrument state
Backwards Compatibility
Notes
In legacy analyzers, only a Reference marker could be Fixed, and it was always Fixed.
Additionally it could not be noved. In the X-Series, any marker can be set to Fixed and can be
moved to any X or Y value.
Initial S/W Revision
Prior to A.02.00
Fixed Marker Y Axis Value
Key Path
Marker, Fixed
Example
:CALC:MARK:MODE FIX sets Marker 1 to Fixed.
Notes
See the description under the Marker key.
Dependencies
You cannot directly set the X or Y value of a Fixed marker which has a marker function turned
on. If an attempt is made to actually adjust it while a Marker Function is on, a warning message
is generated.
State Saved
The marker control mode (Normal, Delta, Fixed, Off) and X and Y Axis values are saved in
instrument state
Default Unit
depends on the current selected Y axis unit
Initial S/W Revision
Prior to A.02.00
Fixed Marker Z Axis Value
The Marker Z position determines which of the 301 traces (0–300) the selected marker is on. It
cannot be set above the maximum trace in the Spectrogram window and, unlike the Marker X
position, will not move off screen in the Spectrogram Window if the storage size is smaller than
the number of traces that can be viewed.
Remote Language Compatibility Measurement Application Reference
683