Agilent Technologies Signal Analyzer N9030a User Manual
Page 1281
6 RLC Swept SA Measurement Front-Panel & SCPI Reference
System
Parameter
Setting
Timebase DAC
Calibrated
Timebase DAC setting
Calibrated value
Auto Align State
Normal (if the instrument is not operating with
default alignment data, Off otherwise)
Auto Align All but RF
Off
Auto Align Alert
Time & Temperature
Key Path
System, Alignments
Mode
All
Example
:SYST:DEF ALIG
Notes
Alignment processing that results as the transition to Auto Alignment Normal will be executed
sequentially; thus *OPC? or *WAI will wait until the alignment processing is complete.
Initial S/W Revision
Prior to A.02.00
Backup or Restore Align Data…
Opens the utility for backing-up or restoring the alignment data.
Alignment data for the instrument resides on the hard drive in a database. Agilent uses high
quality hard drives; however it is highly recommended the alignment data be backed-up to storage
outside of the instrument. Additionally, for customers who use multiple CPU Assemblies or
multiple disk drives, the alignment that pertains to the instrument must be transferred to the
resident hard drive after a CPU or hard drive is replaced. This utility facilitates backing-up and
restoring the alignment data.
This utility allows the operator to navigate to any location of the Windows file system. It is intended that
the operator use a USB memory device or Mapped Network Drive to back up the alignment data to
storage outside of the instrument.
Key Path
System, Alignments
Initial S/W Revision
A.02.00
Remote Language Compatibility Measurement Application Reference
1281