Agilent Technologies Signal Analyzer N9030a User Manual
Page 656
6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Input/Output
State Saved
Saved in Input/Output State
Readback line
1-of-N selection [variable]
Backwards Compatibility
Notes
Prior to A.04.00, OFF was the default functionality except when in the Analog Demod application
or with Tune and Listen, in which case it was DAUDio, and there was no selection menu. So for
backwards compatibility with earlier X-Series firmware versions, Auto (:OUTP:ANAL:AUTO ON)
will duplicate the prior behavior.
The DNWB and SANalyzer parameters, which were legal in PSA but perform no function in the
X-Series, are accepted without error.
Initial S/W Revision
A.04.00
More Information
The table below gives the range for each output.
Analog Out
Nominal Range
exc.
(10%
overrange)
Scale Factor
Notes
Off
0 V
Screen
Video
0 – 1 V open
circuit
10%/division
8566 compatible
Log Video
0 – 1 V
terminated
1/(192.66 dB/V)
dB referenced to mixer level, 1V out for –10 dBm
at the mixer.
Linear Video
0 – 1 V
terminated
100%/V
Linear referenced to Ref Level, 1 V out for RF
envelope at the Ref Level.
Demod
Audio
(varies with analyzer setting)
Auto
Selects the Auto state for the Analog Output menu. In this state, the Analog Output will
automatically be set to the most sensible setting for the current mode or measurement.
If you make a selection manually from the Analog Out menu, this selection will remain in force
until you change it (or re-select Auto), even if you go to a mode or measurement for which the
selected output does not apply.
Key Path
Input/Output, Output Config, Analog Out
Remote Command
OUTPut:ANALog:AUTO OFF|ON|0|1
656
Remote Language Compatibility Measurement Application Reference