Agilent Technologies Signal Analyzer N9030a User Manual
Page 173
3 Programming the Analyzer
STATus Subsystem
Mode
All
Remote
Command
:STATus:QUEStionable:TEMPerature:NTRansition
:STATus:QUEStionable:TEMPerature:NTRansition?
Example
STAT:QUES:TEMP:NTR 1
Reference Oscillator Oven not cold will be reported to the Temperature Summary of the Status Questionable
register.
Preset
0
Min
0
Max
32767
Status
Bits/OPC
dependencies
Sequential command
Initial S/W
Revision
Prior to A.02.00
Questionable Temperature Positive Transition
This command determines which bits in the Questionable Temperature Condition register will set
the corresponding bit in the Questionable Temperature Event register when the condition register
bit has a positive transition (0 to 1).
The variable
Mode
All
Remote
Command
:STATus:QUEStionable:TEMPerature:PTRansition
:STATus:QUEStionable:TEMPerature:PTRansition?
Example
STAT:QUES:TEMP:PTR 1
Reference Oscillator Oven became cold will be reported to the Temperature Summary of the Status
Questionable register.
Preset
32767
Min
0
Max
32767
Status
Bits/OPC
dependencies
Sequential command
Initial S/W
Revision
Prior to A.02.00
Remote Language Compatibility Measurement Application Reference
173