Agilent Technologies Signal Analyzer N9030a User Manual
Page 1416
6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Trigger
Remote Command
:TRIGger[:SEQuence]:RFBurst:SLOPe POSitive|NEGative
:TRIGger[:SEQuence]:RFBurst:SLOPe?
Example
TRIG:RFB:SLOP NEG
Couplings
This same slope is used in the RF Burst selection for the trigger source in the Trigger menu and
for the period timer sync source (in the Trigger menu and in the Gate Source menu).
Preset
POSitive
State Saved
Saved in instrument state
Backwards
Compatibility SCPI
:TRIGger[:SEQuence]:FRAMe:RFBurst:SLOPe
Backwards Compatibility
Notes
The legacy :TRIGger[:SEQuence]:SLOPe command affects the slopes for the VID, LINE, EXT1,
EXT2, and RFB triggers.
Initial S/W Revision
Prior to A.02.00
Trig Delay
Controls a time delay during which the analyzer will wait to begin a sweep after meeting the
trigger criteria. You can use negative delay to pre-trigger the instrument in time domain or FFT,
but not in swept spans.
Key Path
Trigger, RF Burst
Remote Command
:TRIGger[:SEQuence]:RFBurst:DELay
:TRIGger[:SEQuence]:RFBurst:DELay?
:TRIGger[:SEQuence]:RFBurst:DELay:STATe OFF|ON|0|1
:TRIGger[:SEQuence]:RFBurst:DELay:STATe?
Example
TRIG:RFB:DEL:STAT ON
TRIG:RFB:DEL 100 ms
Notes
Video trigger delay may be set to negative values, in time domain, FFT and even swept. It makes
intuitive sense in time domain and works well in FFT mode where the bandwidth of the filter
before the video trigger is about 1.25 span. In swept spans, negative settings of Trig Delay are
treated as a zero setting within the internal hardware and the advisory message "Neg. Trig Delay
unavailable in Swept Mode, zero delay used." is generated when such a delay is set.
Preset
Off, 1.000 us
State Saved
Saved in instrument state
Min
–150 ms
Max
500 ms
Default Unit
s
1416
Remote Language Compatibility Measurement Application Reference