6 RLC Swept SA Measurement Front-Panel & SCPI Reference
View/Display
Key Path
View/Display, Display
Remote Command
:DISPlay:WINDow[1]:TRACe:GRATicule:GRID[:STATe] OFF|ON|0|1
:DISPlay:WINDow[1]:TRACe:GRATicule:GRID[:STATe]?
Example
DISP:WIND:TRAC:GRAT:GRID OFF
Notes
The graticule is the set of horizontal and vertical lines that make up the grid/divisions for the x-
axis and y-axis.
Preset
On
State Saved
Saved in instrument state
Initial S/W Revision
Prior to A.02.00
Display Line
Activates an adjustable horizontal line that is used as a visual reference line. The line’s vertical
position corresponds to its amplitude value. The value of the display line (for example, "–
20.3 dBm") appears above the line itself on the right side of the display in the appropriate font.
The display line can be adjusted using the step keys, knob, or numeric keypad. The unit of the
Display Line is determined by the Y axis unit setting under Amplitude. If more than one window
has a display line, the display line of the selected window is controlled.
If the display line is off the screen, it shows as a line at the top/bottom of the screen with an
arrow pointing up or down. As with all such lines (Pk Thresh, Trigger Level, etc.) it is drawn on
top of all traces.
The display line is unaffected by Auto Couple.
Key Path
View/Display, Display
Remote Command
:DISPlay:WINDow[1]:TRACe:Y:DLINe
:DISPlay:WINDow[1]:TRACe:Y:DLINe?
:DISPlay:WINDow[1]:TRACe:Y:DLINe:STATe OFF|ON|0|1
:DISPlay:WINDow[1]:TRACe:Y:DLINe:STATe?
Example
DISP:WIND:TRAC:Y:DLIN:STAT ON
DISP:WIND:TRAC:Y:DLIN:STAT –32 dBm
Preset
Set the Display Line to Off and –25 dBm on Preset. When the Display Line goes from Off to On, if
it is off screen, set it to either the top or bottom of screen, depending on which direction off
screen it was.
The Display Line's value does not change when it is turned off.
State Saved
Saved in instrument state.
Remote Language Compatibility Measurement Application Reference