6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Source
l
SOURce:POWer:MODE SWEep
is equivalent to :SOURce:POWer:SWEep:STATe
ON
Preset
This is unaffected by “Mode Preset” but is set to FIXed on a “Source Preset” or "Restore
Input/Output Defaults".
Initial S/W Revision
A.06.01
Modified at S/W
Revision
A.10.01
Amptd Offset
Offsets the displayed power of the source in the Amplitude parameter. Using the amplitude offset
allows you to take into account any system losses or gains (for example, due to cable loss),
thereby displaying the actual power delivered to the device under test. See the equations under
the Source, Amplitude, Power Sweep key.
Key Path
Source, Amplitude
Mode
SA
Scope
Mode Global
Remote Command
:SOURce:CORRection:OFFSet
:SOURce:CORRection:OFFSet?
Example
:SOUR:CORR:OFFS 5
Sets the displayed source offset power to 5 dB.
Dependencies
If the requested setting of Amptd Offset causes the calculated external source start or stop
Amplitude to exceed the external source capability, a warning status message is generated,
“Data out of Range; clipped to source max/min”. The Show Source Capabilities and Settings
menu can then be examined to check the source capabilities.
This parameter test and clip is also performed at source acquisition.
Preset
This is unaffected by Mode Preset but is set to 0.00dBm on a Source Preset or Restore
Input/Output Defaults.
State Saved
Part of the Input/Output system, which means it is Loaded and Saved with state.
Min
–1000 dB
Max
+1000 dB
Initial S/W Revision
A.06.01
1128