beautypg.com

Normlize (normalize trace data), Syntax, Legacy products – Agilent Technologies Signal Analyzer N9030a User Manual

Page 416: Description, Syntax legacy products

background image

5  Legacy Command Descriptions
NORMLIZE (Normalize Trace Data)

NORMLIZE (Normalize Trace Data)

Syntax

Legacy Products

8560 series

Description

Activates or de-activates the normalization routine for stimulus-response measurements. This
function subtracts trace B from trace A, offsets the result by the value of the normalized reference
position (NRL) and displays the result in trace A.

Normalization is automatically turned off by an instrument preset (IP), or at power on.

Normalization is not available when using linear mode and is mutually exclusive of other trace
math.

Format

NORMLIZE ON|OFF|1|0

NORMLIZE OA

The OA option only returns the current value to the controller; it does not set the active
function to the normalization state.

NORMLIZE?

Query Data Type

1|0

SCPI Equivalent Commands

:TRACe:COPY TRACE1, TRACE3 (if necessary)

:TRACe[2]:UPDate OFF (blank Trace2, which corresponds to TRB)

:TRACe[2]:DISPlay OFF

:CALCulate:NTData[:STATe] OFF|ON|0|1

:CALCulate:NTData[:STATe]?

416

Remote Language Compatibility Measurement Application Reference