6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Trace/Detector
type letter in the trace annotation panel in the Measurement bar. So in the example below, Traces
3, 4, 5 and 6 have Update Off and Traces 4 and 6 have Display Off.
See
"Trace Update State On/Off" on page 1339
.
See
"Trace Display State On/Off" on page 1340
See
"More Information" on page 1340
.
Key Path
Trace/Detector
Notes
The four states of this 1-of-N actually set two variables, Update and Display, to their four
possible combinations:
•
Trace On: Update and Display both On
•
View: Update Off and Display On
•
Blank: Update Off and Display Off
•
Background: Update On, Display Off
See tables below for detail on the SCPI to control these two variables.
Dependencies
When Signal ID is on, this key is grayed out.
Couplings
Selecting a trace type (Clear Write, Trace Average, Max Hold, Min Hold) for a trace (pressing the
key or sending the equivalent SCPI command) puts the trace in Trace On (Update On and
Display On), even if that trace type was already selected.
Selecting a detector for a trace (pressing the key or sending a [:SENS]:DET:TRAC command) puts
the trace in Trace On (Update On and Display On), even if that detector was already selected.
Selecting a math mode other than Off for a trace (pressing the key or sending the equivalent
SCPI command) puts the trace in Trace On (Update On and Display On), even if that math mode
was already selected.
Loading a trace from a file puts that trace in View regardless of the state it was in when it was
saved; as does being the target of a Copy or a participant in an Exchange.
Initial S/W Revision
Prior to A.02.00
Trace Update State On/Off
Key Path
Trace/Detector
Remote Command
:TRACe[1]|2|...6:UPDate[:STATe] ON|OFF|0|1