6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Sweep/Control
Key Path
Trigger, TV, Standard
Example
TRIG:TV:STAN PAL60
Readback
PAL-N
Initial S/W Revision
Prior to A.02.00
SECAM-L
Sets the TV standard to SECAM-L.
Key Path
Trigger, TV, Standard
Example
TRIG:TV:STAN LSEC
Readback
SECAM-L
Initial S/W Revision
Prior to A.02.00
Control Edge/Level
Sets the method of controlling the gating function from the gating signal.
l
Edge: In Edge triggering, the gate opens (after the Delay) on the selected edge (for example,
positive) of the gate signal and closes on the alternate edge (for example, negative).
l
Level: In Level triggering, the gate opens (after the Delay) when the gate signal has achieved a
certain level and stays open as long as that level is maintained.
Key Path
Sweep/Control, Gate
Remote Command
[:SENSe]:SWEep:EGATe:CONTrol EDGE|LEVel
[:SENSe]:SWEep:EGATe:CONTrol?
Example
SWE:EGAT:CONT EDGE
Dependencies
If the Gate Method is FFT the Control key is grayed out and Edge is selected.
If the Gate Source is TV, Frame or Line, the Control key is grayed out and Edge is selected.
Preset
EDGE
State Saved
Saved in instrument state
Backwards
Compatibility SCPI
[:SENSe]:SWEep:TIME:GATE:TYPE
ESA Compatibility
Initial S/W Revision
Prior to A.02.00
Remote Language Compatibility Measurement Application Reference