6  RLC Swept SA Measurement Front-Panel & SCPI Reference
Meas Setup

Meas Setup

The Meas Setup key opens a menu of softkeys that allow you to control the most important
parameters for the current measurement.

In the Meas Setup menu, you may configure Averaging, by setting the Average Number and the Average
Type.

Key Path

Front-panel key

Initial S/W Revision

Prior to A.02.00

Average/Hold Number

Sets the terminal count number N for Average, Max Hold and Min Hold trace types. This number is an
integral part of how the average trace is calculated. Basically, increasing N results in a smoother
average trace.

See

"More Information" on page 789

.

See

"AVER:CLE command" on page 789

.

Key Path

Meas Setup

Remote Command

[:SENSe]:AVERage:COUNt

[:SENSe]:AVERage:COUNt?

Couplings

Restarting any of these functions (Average, Max Hold or Min Hold) restarts all of them, as there
is only one count.

Preset

100

State Saved

Saved in instrument state

Min

1

Max

10000

Status Bits/OPC
dependencies

See 

"Sweep/Control" on page 1152

for a discussion of the Sweeping, Measuring, Settling and

OPC bits, and the Hi Sweep line. All are affected when a sequence is reset.

Backwards Compatibility
Notes

In the past, when changing the Average Count (now Average/Hold Number), you had to re-
start the trace at the beginning of a sweep to ensure valid average data. Now, the system will
ensure valid results when changing the count limit.