6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Input/Output
Calibrate
Invokes the guided probe calibration. The guided probe calibration is context sensitive and
depends on the channel (I or Q) and the Differential Input state. The calibration is only
performed on the selected channel. When Differential Input is on, both the probe attached to the
main port and the probe attached to the complementary port are calibrated. When Differential
Input is off, only the probe attached to the main port is calibrated. See
.
Key Path
Input/Output, I/Q, I Setup | Q Setup, I Probe | Q Probe, Coupling
Readback Text
The last calibration date, or if no calibration exists, "(empty)".
Last:
Example:
Last: 8/22/2007
1:02:49 PM
Initial S/W Revision
Prior to A.02.00
Next
Perform the Q port calibration.
Key Path
Input/Output, I/Q, Q Setup, Q Probe, Calibrate
Remote Command
:CALibration:IQ:PROBe:Q
Example
CAL:IQ:PROB:Q
Notes
The Q port must be connected to the Cal Out port before issuing the SCPI command.
The calibration data is saved as soon as the port is calibrated and will survive power cycles. It is
not reset by any preset or restore data commands.
State Saved
No
Initial S/W Revision
Prior to A.02.00
Clear Calibration
Clears the calibration data for the current port and probe. It does not clear the data for other
probe types or other ports. If the sensed probe has EEPROM identification, only the data for that
specific probe is cleared. After this command has completed, the probe calibration state will be
the same as if no probe calibration had ever been performed for the specified channel and probe.
The probe attenuation will be the default value for that probe type and the Cable Calibration
frequency response corrections will be used. This command is dependent on the Differential Input