Agilent Technologies Signal Analyzer N9030a User Manual
Page 754
6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Marker Function
•
Each detector is then read in successive single-point zero span sweeps, using a sweep time
equal to the specified dwell time. The value displayed by Measure at Marker represents the
maximum value output by the detector during the dwell timeAutocoupled bandwidth and
average type settings are used for each detector unless the BW & Avg Type key is set to As Set, in
which case the current bandwidth and average type settings are used.
•
Each result is then displayed in the measure at marker window as it becomes available.
•
The analyzer returns to its pre-Measure at Marker span and settings after executing a Measure
at Marker function, including Bandwidth, Avg Type, and EMC Std - regardless of the setting of
BW & Avg Type
.
•
Finally, if the sweep had to be aborted, the aborted sweep is restarted.
While the function is executing, all the fields except Freq show “---“ for their values until the
measurement is complete for that detector. As each detector is read, an informational message is
displayed in the status line, for example:
Measuring with detector 1 (Peak) with RBW=120 kHz
After the last detector, the status line is cleared.
Meas at Marker Window
This key opens a menu which controls the Measure at Marker window.
Key Path
Marker Function, Measure at Marker
Readback
In square brackets, the state of the window then the window position, separated by commas, as
[On, Left]
Initial S/W Revision
A.02.00
Window
This key turns the Measure at Marker window on and off. It turns on automatically when
Measure at Marker is initiated and turns off on a Preset. If the Window is turned on without a
Measure at Marker result, “---“ is displayed for each result for which the detector is not “Off”.
Key Path
Marker Function, Measure at Marker, Meas at Marker Window
Remote Command
:DISPlay:WINDow:MAMarker[:STATe] ON|OFF|1|0
:DISPlay:WINDow:MAMarker[:STATe]?
Example
:DISP:WIND:MAM ON
Couplings
The window turns on automatically when Measure at Marker is initiated and turns off on a
Preset.
754
Remote Language Compatibility Measurement Application Reference