6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Sweep/Control
External 1
Pressing this key, when it is not selected, selects an external input signal as the trigger. A new
sweep/measurement will start when the external trigger condition is met using the external 1
input connector on the rear panel.
Pressing this key, when it is already selected, accesses the external 1 trigger setup menu.
Key Path
Trigger
Example
TRIG:SOUR EXT1 Swept SA measurement
TRIG:
Dependencies
Grayed out if in use by Point Trigger in the Source Setup menu.
Forced to Free Run if already selected and Point Trigger is set to External 1.
State Saved
Saved in instrument state
Status Bits/OPC
dependencies
The Status Operation Register bit 5 "Waiting for Trigger" is set at the same time as the Sweeping
or Measuring bit is set. It is cleared when the trigger actually occurs (that is, after the trigger
event occurs and all the applicable trigger criteria have been met). A corresponding pop-up
message ("Waiting for trigger") is generated if no trigger signal appears after approximately 2
sec. This message goes away when a trigger signal appears.
Initial S/W Revision
Prior to A.02.00
Trigger Level
Sets the value where the external 1 trigger input will trigger a new sweep/measurement.
Key Path
Trigger, External 1
Remote Command
:TRIGger[:SEQuence]:EXTernal1:LEVel
:TRIGger[:SEQuence]:EXTernal1:LEVel?
Example
TRIG:EXT1:LEV 0.4 V
Couplings
This same level is used for the Ext1 trigger source in the Trigger menu, for the Ext1 selection in
the Periodic Timer sync source (in the Trigger menu and in the Gate Source menu), and also for
the Ext1 selection in the Gate Source menu.
Preset
1.2 V
State Saved
Saved in instrument state
Min
–5 V
Max
5 V
Default Unit
V
Remote Language Compatibility Measurement Application Reference