6  RLC Swept SA Measurement Front-Panel & SCPI Reference

Sweep/Control

[:SENSe]:SWEep:EGATe:TIME?

Example

SWE:EGAT:TIME 500 ms

Dependencies

Gate View Sweep Time is initialized:

On Preset (after initializing delay and length).

Every time the Gate Method is set/changed.

Additionally, in the Swept SA measurement, whenever you do a Preset, or leave Gate View, the
analyzer remembers the Gate Delay and Gate Length settings. Then, when returning to Gate
View, if the current Gate Delay and/or Gate Length do not match the remembered values Gate
View Sweep Time is re-initialized.

1. Compute the location of the "gate stop" line, which you know is at time t = tmin + GateDelay

+ GateLength.

Preset

519.3 µs

WiMAX OFDMA: 5 ms

GSM/EDGE: 1 ms

State Saved

Saved in instrument state

Min

1 µs

Max

6000 s

Initial S/W Revision

Prior to A.02.00

Gate View Start Time

Controls the time at the left edge of the Gate View.

Key Path

Sweep/Control, Gate, Gate View Setup

Remote Command

[:SENSe]:SWEep:EGATe:VIEW:STARt

[:SENSe]:SWEep:EGATe:VIEW:STARt?

Example

SWE:EGAT:VIEW:STAR 10ms

Notes

Units of time are required or no units; otherwise an invalid suffix error message will be
generated. See error –131.

Preset

0 ms

State Saved

Saved in instrument state

Min

0

Max

500 ms

Initial S/W Revision

A.10.00

Remote Language Compatibility Measurement Application Reference