6  RLC Swept SA Measurement Front-Panel & SCPI Reference
System

Key Path

Visual annotation in the Show Alignment Statistics screen

Mode

All

Remote Command

:CALibration:TEMPerature:CURRent?

Example

:CAL:TEMP:CURR?

Notes

Value is in degrees Centigrade.

Value is invalid if using default alignment data (Align Now, All required)

State Saved

No

Initial S/W Revision

Prior to A.02.00

Key Path

Visual annotation in the Show Alignment Statistics screen

Mode

All

Remote Command

:CALibration:TIME:LALL?

Example

:CAL:TIME:LALL?

Notes

Value is the elapsed time, in seconds, since the last successful Align Now, All or Align Now, All
but RF was executed.

State Saved

No

Initial S/W Revision

Prior to A.02.00

Key Path

Visual annotation in the Show Alignment Statistics screen

Mode

All

Remote Command

:CALibration:TEMPerature:LALL?

Example

:CAL:TEMP:LALL?

Notes

Value is in degrees Centigrade at which the last successful Align Now, All or Align Now, All but
RF was executed.

State Saved

No

Initial S/W Revision

Prior to A.02.00

Key Path

Visual annotation in the Show Alignment Statistics screen

Mode

All

Remote Command

:CALibration:TIME:LRF?