beautypg.com

Ksx (external trigger), Syntax, Legacy products – Agilent Technologies Signal Analyzer N9030a User Manual

Page 349: Description, Syntax legacy products

background image

5  Legacy Command Descriptions

KSx (External Trigger)

KSx (External Trigger)

Syntax

Legacy Products

8566A/B, 8568A/B

Description

Activates the normal external trigger mode. When KSx is executed, the RF input signal is only
displayed when the external trigger level exceeds the trigger threshold level.

Format

KSx

Query Data Type

N/A

SCPI Equivalent Commands

None

Notes

The functions of KSx are identical to TM EXT. See

"TM (Trigger Mode)" on page 489

.

If an 8566A/B or an 8568A/B analyzer is in zero span and the sweep time is less than 20
msec, the display is refreshed only when a fresh trace has been taken. This can cause the
displayed trace to flicker.

In X-Series instruments, all traces are displayed continuously, so are therefore free of
flicker.

Remote Language Compatibility Measurement Application Reference

349