6  RLC Swept SA Measurement Front-Panel & SCPI Reference

Trigger

Trigger Level

Sets the value where the external 2 trigger input will trigger a new sweep/measurement.

Key Path

Trigger, External 2

Remote Command

:TRIGger[:SEQuence]:EXTernal2:LEVel

:TRIGger[:SEQuence]:EXTernal2:LEVel?

Example

TRIG:EXT2:LEV 1.1 V

Couplings

This same level is used for the Ext2 trigger source in the Trigger menu, for the Ext2 selection in
the Periodic Timer sync source (in the Trigger menu and in the Gate Source menu), and also for
the Ext2 selection in the Gate Source menu.

Preset

1.2 V

State Saved

Saved in instrument state

Min

–5 V

Max

5 V

Default Unit

V

Backwards
Compatibility SCPI

:TRIGger[:SEQuence]:FRAMe:EXTernal2:LEVel

Initial S/W Revision

Prior to A.02.00

Trig Slope

Controls the trigger polarity. It is set positive to trigger on a rising edge and negative to trigger on
a falling edge.

Key Path

Trigger, External 2

Remote Command

:TRIGger[:SEQuence]:EXTernal2:SLOPe POSitive|NEGative

:TRIGger[:SEQuence]:EXTernal2:SLOPe?

Example

TRIG:EXT2:SLOP NEG

Couplings

This same slope is used in the Ext2 selection for the trigger source in the Trigger menu and for
the period timer sync source (in the Trigger menu and in the Gate Source menu).

Preset

POSitive

State Saved

Saved in instrument state

Backwards
Compatibility SCPI

:TRIGger[:SEQuence]:FRAMe:EXTernal2:SLOPe

Backwards Compatibility
Notes

The legacy :TRIGger[:SEQuence]:SLOPe command affects the slopes for the VID, LINE, EXT1,
EXT2, and RFB triggers.

Initial S/W Revision

Prior to A.02.00

Remote Language Compatibility Measurement Application Reference