6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Marker
This flowchart makes it clear that putting all lower-numbered traces in View is the simplest way to
specify which trace you want the markers to go to when they turn on. For example, if you want all
Markers to go to trace 2 when they turn on, put trace 1 in View.
Auto Init OFF
This command associates the marker with the specified trace and turns Marker Trace, Auto Init
OFF for that marker. If the marker is not Off it moves the marker from the trace it was on to the
new trace. If the marker is Off it stays off but is now associated with the specified trace.
The query returns the number of the trace on which the marker is currently placed, even if that
marker is in Auto mode.
Remote Command
:CALCulate:MARKer[1]|2|...12:TRACe:AUTO OFF|ON|0|1
:CALCulate:MARKer[1]|2|...12:TRACe:AUTO?
Notes
Turning Marker Trace Auto Init off has no effect on the trace on which the marker is currently
placed.
The response to the query will be 0 if OFF, 1 if ON.
Couplings
The state of Auto Init is not affected by the Auto Couple key.
Auto Init is set to True on a Preset or All Markers Off.
708
Remote Language Compatibility Measurement Application Reference