Agilent Technologies Signal Analyzer N9030a User Manual
Page 1236
6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Sweep/Control
This command sets gate parameter values to the ESA preset values, as follows:
l
Gate trigger type = edge
l
Gate polarity = positive
l
Gate delay = 1 us
l
Gate length = 1 us
Remote Command
[:SENSe]:SWEep:TIME:GATE:PRESet
ESA Compatibility
Initial S/W Revision
Prior to A.02.00
Gate Level (Remote Command Only)
Sets the gate input transition point level for the external TRIGGER inputs on the front and rear
panel. This is a legacy command for PSA compatibility. It is simply an alias to the equivalent
trigger level command.
Remote Command
[:SENSe]:SWEep:EGATe:EXTernal[1]|2:LEVel
[:SENSe]:SWEep:EGATe:EXTernal[1]|2:LEVel?
Notes
This command is simply an alias to
:TRIGger[:SEQuence]:EXTernal[1]|2:LEVel
For details refer
Initial S/W Revision
Prior to A.02.00
Gate Polarity (Remote Command Only)
Sets the polarity for the gate signal. This setup is now done using the gate trigger’s slope setting.
l
When Positive (Pos) is selected, a positive-going edge (Edge) or a high voltage (Level) will
satisfy the gate condition, after the delay set with the Gate Delay key.
l
When Negative (Neg) is selected, a negative-going edge (Edge) or a low voltage (Level) will
satisfy the gate condition after the delay.
Remote Command
[:SENSe]:SWEep:EGATe:POLarity NEGative|POSitive
[:SENSe]:SWEep:EGATe:POLarity?
Example
SWE:EGAT:POL NEG
1236
Remote Language Compatibility Measurement Application Reference