Mkr->cf step, Mkr->start – Agilent Technologies Signal Analyzer N9030a User Manual
Page 779
6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Marker To
Mkr->CF Step
Sets the center frequency (CF) step size of the analyzer to the marker frequency, or in a delta-
marker mode, to the frequency difference between the delta and reference markers.
If the currently selected marker is not on when this key is pressed, it will be turned on at the
center of the screen as a normal type marker.
Key Path
Marker ->
Remote Command
:CALCulate:MARKer[1]|2|...12[:SET]:STEP
Example
CALC:MARK1:STEP sets the CF step to the value (or delta value) of marker 1.
Notes
Sending this command selects the subopcoded marker
If specified marker is off, this command will turn it on at the center of the screen as a normal
type marker.
Dependencies
This function is not available (key is grayed out) when x-axis is the time domain
Couplings
All the usual couplings associated with setting CF Step apply.
Initial S/W Revision
Prior to A.02.00
Mkr->Start
Changes the start frequency to the frequency of the selected marker. The marker stays at this
frequency, so it moves to the left edge of the display. In delta marker mode, this function sets the
start frequency to the x-axis value of the delta marker.
If the currently selected marker is not on when this key is pressed, it will be turned on at the
center of the screen as a normal type marker.
Key Path
Marker ->
Remote Command
:CALCulate:MARKer[1]|2|...12[:SET]:STARt
Example
CALC:MARK1:STAR sets the start frequency to the value (or delta value) of marker 1.
Notes
Sending this command selects the subopcoded marker
If specified marker is off, this command will turn it on at the center of the screen as a normal
type marker.
Dependencies
This function is not available (key is grayed out) when x-axis is the time domain
Couplings
All the usual couplings associated with setting Start Frequency apply.
Initial S/W Revision
Prior to A.02.00
Remote Language Compatibility Measurement Application Reference