3 Programming the Analyzer
Remote Measurement Functions
Non-Constant Envelope
See table below for explanation of variables.
Variable
Description
Start Offset is an optional real number.
Its unit is seconds for time-domain traces, or a dimensionless index (0 to Npoints – 1) for frequency-domain
traces.
It specifies the amount of data at the beginning of the trace that will be ignored before the decimation process
starts. It is the time or frequency change from the start of the trace to the point where you want to start using
the data.
The default value is zero.
An optional real number.
Its unit is seconds for time-domain traces, or a dimensionless index (0 to Npoints – 1) for frequency-domain
traces.
It defines how much data will be compressed into one value.
This parameter has a default value equal to the current trace length.
Repeat Offset is an optional real number.
Remote Language Compatibility Measurement Application Reference