5 Legacy Command Descriptions
NRPOS (Normalized Reference Position)
NRPOS (Normalized Reference Position)
Syntax
Legacy Products
8560 series
Description
Adjusts the normalized reference-position that corresponds to the position on the graticule where
the difference between the measured and calibrated traces reside. The dB value of the normalized
reference position is equal to the normalized reference level. The normalized reference position
can be adjusted between 0.0 and 10.0, corresponding to the bottom and top graticule lines,
respectively.
Format
NRPOS
Range: Min = 0; Max = 10
NRPOS UP|DN
UP or DN increments by 1.0
NRPOS OA
NRPOS?
Query Data Type
Returns the current Normalized Reference Position.
SCPI Equivalent Commands
:DISPlay:WINDow[1]:TRACe:Y[:SCALe]:NRPosition
)
Preset
10