Agilent Technologies Signal Analyzer N9030a User Manual
Page 1408
6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Trigger
Example
TRIG:EXT1:LEV 0.4 V
Couplings
This same level is used for the Ext1 trigger source in the Trigger menu, for the Ext1 selection in
the Periodic Timer sync source (in the Trigger menu and in the Gate Source menu), and also for
the Ext1 selection in the Gate Source menu.
Preset
1.2 V
State Saved
Saved in instrument state
Min
–5 V
Max
5 V
Default Unit
V
Backwards
Compatibility SCPI
:TRIGger[:SEQuence]:EXTernal:LEVel
For backward compatibility, the parameter EXTernal is mapped to EXTernal1
Backwards
Compatibility SCPI
:TRIGger[:SEQuence]:FRAMe:EXTernal1:LEVel
Initial S/W Revision
Prior to A.02.00
Trig Slope
Controls the trigger polarity. It is set positive to trigger on a rising edge and negative to trigger on
a falling edge.
Key Path
Trigger, External 1
Remote Command
:TRIGger[:SEQuence]:EXTernal1:SLOPe POSitive|NEGative
:TRIGger[:SEQuence]:EXTernal1:SLOPe?
Example
TRIG:EXT1:SLOP NEG
Couplings
This same slope is used in the Ext1 selection for the trigger source in the Trigger menu and for
the period timer sync source (in the Trigger menu and in the Gate Source menu).
Preset
POSitive
State Saved
Saved in instrument state
Backwards
Compatibility SCPI
:TRIGger[:SEQuence]:EXTernal:SLOPe
For backward compatibility, the parameter EXTernal is mapped to EXTernal1
Backwards
Compatibility SCPI
:TRIGger[:SEQuence]:FRAMe:EXTernal1:SLOPe
Backwards Compatibility
Notes
The legacy :TRIGger[:SEQuence]:SLOPe command affects the slopes for the VID, LINE, EXT1,
EXT2, and RFB triggers.
Initial S/W Revision
Prior to A.02.00
1408